DC1-300

Immersion Dip Cooler (ambient to -20°C)

Cena na dotaz

0 Kč Vč. DPH

0 Kč bez DPH

Dimensions
External: 345w x 445d x265h mm
Packed: 45 x 41 x 32 cm
Cooling coil: Stainless steel coil wound diameter 65mmOn the end of a 900mm long flexible cooling line.
Temperature Range: -20 to 20°CRefer to Recommended fluids for fluid type and working temperature range
Temperature control: None. Continuously providing counter cooling with water bath providing temperature control
Cooling power:
10°C: 800W
0°C: 580W
-10°C: 400W
-20°C: 250W
Refrigerant: R134a
Compressor: Danfoss
Voltage 230V 50-60Hz
Packed Weight: 22 kg
 
Description
This Clifton® Immersion Dip refrigeration cooler has no temperature control, being easy to use makes this model an ideal instruments for a variety of cooling applications, suiting an immersion cooling coil. The immersion cooler runs continuously providing counter cooling with bath providing temperature control.

Immersion Coolers replaces wasteful use of cold water cooling running to waste are quiet in operation and can be used on bench top.

There are no controls, Immersion Cooler operates continually counter cooling a Stirred or Circulating bath, control is provided by water bath to maintain set temperature.

Featuring a flexible 900mm long stainless steel pipe with a protective outer shield, exiting from top of the unit, making the stainless steel cooling coil naturally stand in the corner of any Clifton NE4 Stirred or Circulating water bath. This maximises the working area within the bath and reduces the chance of ‘cold shock’, a chance of the coil freezing cold, accidentally intruding or touching and spoiling work.


Heat-up Times
  • 0°C - 60 mins
  • -15°C - 120 mins
  • -20°C - 180 mins
Recommended fluids:
50% water 50% glycol: -20°C to 20°C
Lab Bath 4590: -15°C to 20°C
Heating/Cooling Information:
Recommend operating in ambient temperature 20°C.

In use we recommend using the lid at all times reduces evaporation and also assists keeping water free from contamination and saves energy reducing electricity consumption and running costs.

Power Consumption (220-240V): 600W
Warranty Period:3 Years
Notes
Normal operating performance, working ambient temperature range 10°C to 25°C. Where ambient temperature range exceeds 20°C this degrades cooling performance. For optimum cooling performance reduce the effect of high ambient temperatures. Use with a minimum distance of 40cm of free air around ventilation slots - DO NOT obstruct. The Immersion Dip Cooler is air cooled with air from its surroundings sucked in at the front and warm air expelled out of the rear. To provide good air circulation and operating performance keep the ventilation slots free from obstruction. To achieve optimum performance with any Stirred or Circulating bath should always be covered with a Lid or BP0368 Polypropylene spheres at all times reducing up to 70% of the heat, from ambient room temperatures, transferring into the bath being cooled. Doubling this layer of polypropylene spheres or using a Lid and spheres reduces positive heating up to 90%. DO Not use with Unstirred water baths otherwise the cooling coil will freeze solid and cooling performance lost. Keep out of direct sunlight, away from high ambient temperatures or direct heat sources.Do not place it where there are corrosive fumes, excessive moisture or excessively dusty areas.